Relative stopping power measurements and prosthesis artifacts reduction in proton CT / Civinini, C.; Scaringella, M.; Brianzi, M.; Intravaia, M.; Randazzo, N.; Sipala, V.; Rovituso, M.; Tommasino, F.; Schwarz, M.; Bruzzi, M.. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - 65:22(2020), p. 225012. [10.1088/1361-6560/abb0c8]
Abstract:
We present a set-up for proton computed tomography (pCT), composed of a microstrip silicon tracker and a YAG:Ce calorimeter, able to directly measure the relative stopping power (RSP) maps to be used in hadron therapy. The system, tested with an electron density phantom at the Trento proton Therapy Center, is able to correlate measured and expected RSP with discrepancies of the order of 1% or less. Furthermore, pCT tomographies of an anthropomorphous head phantom taken with our device, when compared with X-ray CT images of the same object, evidence a significant reduction of artifacts induced by titanium spinal bone prosthesis and tungsten dental filling.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Algebraic reconstruction techniques; Hadron therapy; Metallic prosthesis; Proton tomography; Stopping power; Calibration; Head; Humans; Image Processing, Computer-Assisted; Phantoms, Imaging; Artifacts; Prostheses and Implants; Protons; Tomography, X-Ray Computed