Assessment and validation of methods for the determination of γ-glutamyltransferase activity in sheep milk
Articolo
Data di Pubblicazione:
2009
Citazione:
Assessment and validation of methods for the determination of γ-glutamyltransferase activity in sheep milk / Piga, C; Urgeghe, Pietro Paolo; Piredda, G; Scintu, Mf; Sanna, Gavino. - In: FOOD CHEMISTRY. - ISSN 0308-8146. - 115:4(2009), pp. 1519-1523. [10.1016/j.foodchem.2009.01.041]
Abstract:
The determination of enzymatic activity in sheep milk is still today a practically unexplored field of
research. This study proposes and fully validates two analytical procedures (i.e. by means of UV–vis spectrophotometry
and RP-HPLC methods) for the determination of c-glutamyltransferase activity in sheep
milk. Both methods are characterised by low detection and quantification limits, excellent linearity over
a wide enzymatic activity interval, very good repeatability and reproducibility, and are bias-free. The RPHPLC
method provides better sensitivity and automation capability levels than that of UV–vis, and its use
is hence suggested for screening purposes. These methods have been preliminarily tested with a number
of real samples of whole sheep milk, obtaining an average c-glutamyltransferase activity value of
2.93 ± 0.50 U ml1 and a range from 2.72 ± 0.10 U ml1 to 3.46 ± 0.11 U ml1.
research. This study proposes and fully validates two analytical procedures (i.e. by means of UV–vis spectrophotometry
and RP-HPLC methods) for the determination of c-glutamyltransferase activity in sheep
milk. Both methods are characterised by low detection and quantification limits, excellent linearity over
a wide enzymatic activity interval, very good repeatability and reproducibility, and are bias-free. The RPHPLC
method provides better sensitivity and automation capability levels than that of UV–vis, and its use
is hence suggested for screening purposes. These methods have been preliminarily tested with a number
of real samples of whole sheep milk, obtaining an average c-glutamyltransferase activity value of
2.93 ± 0.50 U ml1 and a range from 2.72 ± 0.10 U ml1 to 3.46 ± 0.11 U ml1.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Glutamyltransferase; Enzymatic activity; Sheep milk
Elenco autori:
Piga, C; Urgeghe, Pietro Paolo; Piredda, G; Scintu, Mf; Sanna, Gavino
Link alla scheda completa:
Pubblicato in: