Data di Pubblicazione:
2005
Citazione:
Prototype tracking studies for proton CT / Feldt, F; Heimann, J; Blumenkrantz, N; Lucia, D; Sadrozinski, H. F. W.; Seiden, A; Sowerwine, W; Williams, D. C.; Bashkirov, V; Schulte, R; Bruzzi, M; Menichelli, D; Scaringella, M; Cirrone, G. A. P.; Cuttone, G; Randazzo, N; Sipala, Valeria; LO PRESTI, D.. - 1:(2005), pp. 59-63. (Intervento presentato al convegno Nuclear Science Symposium and Medical Imaging Conference, IEEE 2005 tenutosi a Fajardo nel 23-29 Oct. 2005) [10.1109/NSSMIC.2005.1596207].
Abstract:
As part of a program to investigate the feasibility of proton computed tomography, the most likely path (MLP) of protons inside an absorber was measured in a beam experiment using a silicon strip detector set-up with high position and angular resolution. The locations of 200 MeV protons were measured at three different absorber depth of PMMA (3.75, 6.25 and 12.5 cm) and binned in terms of the displacement and exit angle measured behind the absorber. The observed position distributions were compared to theoretical predictions showing that the location of the protons can be predicted with an accuracy of better than 0.5 mm
Tipologia CRIS:
4.1 Contributo in Atti di convegno
Elenco autori:
Feldt, F; Heimann, J; Blumenkrantz, N; Lucia, D; Sadrozinski, H. F. W.; Seiden, A; Sowerwine, W; Williams, D. C.; Bashkirov, V; Schulte, R; Bruzzi, M; Menichelli, D; Scaringella, M; Cirrone, G. A. P.; Cuttone, G; Randazzo, N; Sipala, Valeria; LO PRESTI, D.
Link alla scheda completa:
Titolo del libro:
Nuclear Science Symposium Conference Record, 2005 IEEE
Pubblicato in: