Publication Date:
2002
Short description:
Experimental study of Compton scattering reduction in digital mammographic imaging / Amendolia, Salvator Roberto; Oliva, Piernicola; Quattrocchi, Mariagrazia; Rosso, Valeria; Stefanini, Arnaldo; Zucca, Sergio; Bisogni, Maria Giuseppina; Novelli, Marzia; Delogu, Pasquale; Fantacci, Maria Evelina. - In: IEEE TRANSACTIONS ON NUCLEAR SCIENCE. - ISSN 0018-9499. - 49:5, Part 1(2002), pp. 2361-2365. [10.1109/TNS.2002.803816]
abstract:
In mammography, the first cause of image contrast reduction arises from the photons scattered inside the examined organ. The amount of Compton scattering strongly depends on the irradiation area and on the distance between the organ and the X-ray detector. We have experimentally evaluated how these geometrical conditions affect the scattering fraction. Our experimental setup includes a single photon counting device based on a silicon pixel detector as X-ray sensor; a lucite cylinder to simulate the breast tissue, and a lead collimator to define the irradiation area. We have evaluated the contrast and the signal-to-noise ratio for images acquired in different conditions.
Iris type:
1.1 Articolo in rivista
Keywords:
Contrast; mammography; pixel detector; scattered
photons; signal-to-noise ratio (SNR); single photon counting
chip
List of contributors:
Amendolia, Salvator Roberto; Oliva, Piernicola; Quattrocchi, Mariagrazia; Rosso, Valeria; Stefanini, Arnaldo; Zucca, Sergio; Bisogni, Maria Giuseppina; Novelli, Marzia; Delogu, Pasquale; Fantacci, Maria Evelina
Published in: