Performance of a SiPM-based readout of a large CsI(Tl) crystal irradiated with low energy protons
Conference Paper
Publication Date:
2017
Short description:
Performance of a SiPM-based readout of a large CsI(Tl) crystal irradiated with low energy protons / Bondi, M.; Battaglien, M.; Carpinelli, M.; Celentano, A.; De Napoli, M.; De Vita, R.; Leonora, E.; Marsicano, L.; Randazzo, N.; Sipala, V.. - 2017-:(2017), pp. 1-2. ( 2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016 fra 2016) [10.1109/NSSMIC.2016.8069804].
abstract:
In the present work we report a study of the light response of a CsI(Tl) crystal readout by a SiPM to low energy protons, accelerated by the Tandem Van der Graaf at the Laboratori Nazionali del Sud (LNS)-INFN, Catania, Italy. The light yield was measured for protons of different kinetic energies, from 2.5 MeV to 24 MeV. We also derived the light attenuation length. The results of this work indicate that a CsI coupled with a SiPM-based modern readout solution allows for the detection of low energy signals.
Iris type:
4.1 Contributo in Atti di convegno
Keywords:
Radiology, Nuclear Medicine and Imaging; Instrumentation; Nuclear and High Energy Physics; Electronic, Optical and Magnetic Materials
List of contributors:
Bondi, M.; Battaglien, M.; Carpinelli, M.; Celentano, A.; De Napoli, M.; De Vita, R.; Leonora, E.; Marsicano, L.; Randazzo, N.; Sipala, V.
Book title:
2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop, NSS/MIC/RTSD 2016